User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Keithley 130: (1 Year) CAL VER DATE: 07-Sep-94 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 43 NUMBER OF LINES: 151 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON #! Note: This procedure was extracted from the specification section of the #! Keithley 130 manual and adjusted to provide all the check points #! used by Keithley for prove in. # # 1.001 ASK+ B X 1.002 ASK- R N P J F W 1.003 MMFC C2 1.004 5700 * S # # 1.005 HEAD [32] [27][91]1mEQUIPMENT SETUP 1.006 DISP WARNING !! HIGH VOLTAGE is used or exposed during the 1.006 DISP performance of this calibration. DEATH ON CONTACT may 1.006 DISP result if personnel fail to observe safety precautions. 1.006 DISP 1.006 DISP Ambient temperature: 23øC ñ5øC 1.006 DISP [32] Relative humidity: ó80% # # 1.007 HEAD [32] [27][91]1mCONNECTION DATA 1.008 DISP Connect the 5700A and the 130 as follows. 1.008 DISP [32] 5700A 130 1.008 DISP [32] OUTPUT VêA HI ÄÄÄÄÄÄÄÄÄÄ>Â> V-ê 1.008 DISP [32] SENSE Vê HI ÄÄÄÄÄÄÄÄÄÄÄÙ 1.008 DISP [32] OUTPUT VêA LO ÄÄÄÄÄÄÄÄÄÄ>Â> COM 1.008 DISP [32] SENSE Vê LO ÄÄÄÄÄÄÄÄÄÄÄÙ 1.008 DISP AUX CURRENT OUTPUT HI ÄÄÄÄÄÄÄÄÄÄÄÄ> mA # 1.008 DISP Connect the 5725A and the 130 as follows. 1.009 DISP [32] 5725A UUT 1.009 DISP [32] CURRENT OUTPUT HI ÄÄÄÄÄÄÄÄÄÄÄÄ> 10A 1.009 DISP [32] CURRENT OUTPUT LO ÄÄÄÄÄÄÄÄÄÄÄÄ> 10A COM 1.010 JMP 2.001 1.011 EVAL Dummy # # 2.001 HEAD [27][91]1m{DC Volts Verification} 2.002 DISP 1. Turn on the Keithley 130 Digital Multimeter 2.002 DISP 2. Select the DCV function and the 200mV range.[7] 2.003 5700 0mV 0.5% 0.1U 2W 3.001 5700 190mV 0.5% 0.1U 2W 4.001 5700 -190mV 0.5% 0.1U 2W # 5.001 DISP Select the 2 Volt range.[7] 5.002 5700 1.9V 0.5% 0.001U 2W # 6.001 DISP Select the 20 Volt range.[7] 6.002 5700 19V 0.5% 0.01U 2W # 7.001 DISP Select the 200 Volt range.[7] 7.002 5700 155.7V 0.5% 0.1U 2W 8.001 5700 190V 0.5% 0.1U 2W # 9.001 DISP Select the 1000 Volt range.[7] 9.002 5700 1000V 0.5% 1U 2W # # 10.001 HEAD [27][91]1m{AC Volts Verification} 10.002 DISP 2. Select the ACV function.[7] 10.003 5700 750V 1% 5U 50H 2W 11.001 5700 750V 1% 5U 60H 2W 12.001 5700 750V 1% 5U 100H 2W # 13.001 DISP Select the 200 Volt range.[7] 13.002 5700 190V 1% 0.5U 50H 2W 14.001 5700 190V 1% 0.5U 60H 2W # 15.001 DISP Select the 20 Volt range.[7] 15.002 5700 13.28V 1% 0.05U 500H 2W 16.001 5700 19V 1% 0.05U 50H 2W 17.001 5700 19V 1% 0.05U 60H 2W # 18.001 DISP Select the 2 Volt range.[7] 18.002 5700 1.9V 1% 0.005U 50H 2W 19.001 5700 1.9V 1% 0.005U 60H 2W 20.001 5700 1.9V 1% 0.005U 100H 2W 21.001 5700 1.9V 1% 0.005U 500H 2W # 22.001 DISP Select the 200 mVolt range.[7] 22.002 5700 190mV 1% 0.5U 50H 2W 23.001 5700 190mV 1% 0.5U 60H 2W 24.001 5700 190mV 1% 0.5U 500H 2W # # 25.001 HEAD [27][91]1m{DC Amps Verification} 25.002 DISP 2. Select the DCA function and the 2mA range.[7] 25.003 5700 1.9mA 1% 0.001U 2W # 26.001 DISP Select the 20 mA range.[7] 26.002 5700 -12.35mA 1% 0.01U 2W 27.001 5700 19mA 1% 0.01U 2W 28.001 5700 10A 2% 0.01U B1 2W # 29.001 DISP Select the 200 mA range.[7] 29.002 5700 190mA 1% 0.1U 2W # 30.001 DISP Select the 2000 mA range.[7] 30.002 5700 1900mA 2% 1U 2W # # 31.001 HEAD [27][91]1m{AC Amps Verification} 31.002 DISP 2. Select the ACA function.[7] 31.003 5700 1000mA 3% 5U 500H 2W 32.001 5700 1900mA 3% 5U 60H 2W # 33.001 DISP Select the 200 mA range.[7] 33.002 5700 190mA 2% 0.5U 60H 2W 34.001 5700 190mA 2% 0.5U 500H 2W # 35.001 DISP Select the 20 mA range.[7] 35.002 5700 19mA 2% 0.05U 60H 2W 36.001 5700 10A 3% 0.05U 60H B1 2W # 37.001 DISP Select the 2 mA range.[7] 37.002 5700 1.9mA 2% 0.005U 60H 2W # # 38.001 HEAD [27][91]1m{Ohms Verification} 38.002 DISP 2. Select the ês function and the 200ê range.[7] 38.003 5700 100Z 0.5% 0.4U CW # 39.001 DISP Select the 2Kê range.[7] 39.002 5700 1.0kZ 0.5% 0.001U CW # 40.001 DISP Select the 20Kê range.[7] 40.002 5700 10kZ 0.5% 0.01U CW # 41.001 DISP Select the 200Kê range.[7] 41.002 5700 100kZ 0.5% 0.1U 2W # 42.001 DISP Select the 20Mê range.[7] 42.002 5700 10MZ 2% 1U 2W # # 43.001 HEAD [27][91]1mKeithley 130: (1 Year) CAL VER Completed 43.002 5700 * S 43.003 END